Official recall record
Impella® Controller (ref 77714)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-07-09
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
Abiomed has identified an Automated Impella Controller™ (AIC) issue that may impair the detection of an Impella® pump when connected to an AIC. The pump detection issue may occur with any of the Abiomed Impella® pumps and may occur during console-to-console transfer or case start. There is no visual alarm displayed on the AIC screen to indicate the detection issue in these situations. Recall start date: July 7, 2025