Official recall record
Osstell® Smartpeg Type 101-104
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-07-17
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
If not correctly used with a reference measurement on implant level. The measurement on abutment can indicate a false high ISQ value. This can influence the clinician to take the wrong decision and load the implant too early, in the extension potentially resulting in implant failure for the patient. Recall start date: May 23, 2025
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.