epoc® Blood Analysis BGEM Test Cards
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-07-18
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in some epoc® test card lots. All other analytes are performing as intended. The observed average bias for sodium was -4.4 mmol/l. The maximum bias observed was -14 mmol/l, which occurred at a higher sodium concentration around 150 mmol/l. No positive bias has been observed. Quality control (qc) fluid may or may not detect the issue depending on the bias magnitude and the frequency of qc testing. Recall start date: July 9, 2025