ProductGuru Recalls

Official recall record

Philips Achieva and Ingenia Systems

Health Canada · 2026-07-20 · CanadaType II
Issuing authority
Health Canada
Alert date
2026-07-20
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Philips has identified an issue where component failures in the gradient coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire. If a component failure occurs, the user may observe the following: repeated scan abort error messages due to gradient amplifier fault detection unusual noise from the system in the examination or technical room burning smell smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or death. This issue could also lead to property damage. The scope includes MR systems with gradient coil types WA15, WA30S, WB30S, and AG30S. Recall start date: July 7, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — official record →
Record last checked against the registry on 2026-07-29.

More recalls: Health Canada · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.