Official recall record
Philips Achieva and Ingenia Systems
- Issuing authority
- Health Canada
- Alert date
- 2026-07-20
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Philips has identified an issue where component failures in the gradient coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire. If a component failure occurs, the user may observe the following: repeated scan abort error messages due to gradient amplifier fault detection unusual noise from the system in the examination or technical room burning smell smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may include inhalation of smoke, burns, and/or asphyxia which may lead to injury or death. This issue could also lead to property damage. The scope includes MR systems with gradient coil types WA15, WA30S, WB30S, and AG30S. Recall start date: July 7, 2025
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.