ProductGuru Recalls

Official recall record

Unicel Dxi 600 Access Immunoassay System- Thyroglobulin Assay

Health Canada — Medical Devices · 2025-07-23 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-07-23
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Multiple customers reported receiving high patient results when using access thyroglobulin reagent pack lot 439163. An internal review of batch records determined this was the first lot released using antibodies produced by BEC instead of a third party supplier. Recall start date: July 2, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.