Corin Operating Tables
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-07-24
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Under certain conditions - such as the simultaneous use of the operating table with other medical devices that may cause interference, including High-Frequency (HF) surgical equipment - we have identified a potential issue that could prevent the device from performing as intended. If the issue occurs, the Corin operating table does not respond to any command. It is impossible to actuate any movement of the Corin operating table, neither with the universal remote control nor the override panel. We are not aware of instances where a patient was injured due to the described issue. Recall start date: July 18, 2025