ProductGuru Recalls

Official recall record

VITROS Chemistry Product System-Ca Slides (Calcium)

Health Canada — Medical Devices · 2025-07-24 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-07-24
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Generations (Gens) 67 and above of VITROS CA slides may experience an increased occurrence of condition code th4-63j during calibration, on VITROS XT 3400 and VITROS XT 7600 systems. Condition code th4-63j is generated when the reflectometer detects image variations which exceed an assay parameter defined on the assay data disk (add). If condition code th4-63j is generated during calibration, the calibration may fail. Recall start date: July 14, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.