Official recall record
VITROS Chemistry Product System-Ca Slides (Calcium)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-07-24
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Generations (Gens) 67 and above of VITROS CA slides may experience an increased occurrence of condition code th4-63j during calibration, on VITROS XT 3400 and VITROS XT 7600 systems. Condition code th4-63j is generated when the reflectometer detects image variations which exceed an assay parameter defined on the assay data disk (add). If condition code th4-63j is generated during calibration, the calibration may fail. Recall start date: July 14, 2025