Official recall record
STA – Liquid Anti-Xa
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-07-25
- Last updated (per notice)
- 2025-07-31
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Following internal tests, contamination of STA®-LIQUID ANTI-Xa by Stago Fibrinogen reagents was identified in UFH and LMWH applications on STA SATELLITE® or STA SATELLITE MAX® analyzers. Contamination occurs when one or more tests with the STA®-Fibrinogen 5 reagent are performed before the STA®-LIQUID ANTI-Xa test for heparin monitoring on an STA SATELLITE® or STA SATELLITE MAX® analyzer. This contamination causes an underestimation of heparin levels which could lead to a heparin treatment change and a subsequent overdose risk for the patient. Recall start date: July 21, 2025
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.