What the authority reported

Baxter Corporation is issuing an urgent medical device correction for Centrella Smart+ beds with system software version 1.27.000 due to the potential for an interface issue with the Rauland Responder 5 nurse call system. The bed exit system will alert at the bed but may fail to send a remote alert through the nurse call system if a remote alert was previously sent and cancelled at the in-room nurse call wall unit. If the bed exit alert fails to notify the caregiver at the assigned remote location (e.g., nurse's station) the staff may be unaware the patient exited the bed if they are unable to hear the local bed alert. As a result, a patient may experience a fall resulting in a critical musculoskeletal injury, including head trauma. To date, Baxter has not received any reports of patient injury related to this issue. Recall start date: September 19, 2024

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.