ProductGuru Recalls

Official recall record

SmartDrive MX2+

Health Canada — Medical Devices · 2025-08-13 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-08-13
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Max Mobility has initiated a voluntary correction of the speed control dial, a component utilized with the SmartDrive MX2+ Power Assist device (this attaches to parent device manual wheelchairs to provide power assist). Through market feedback and subsequent investigation, Max Mobility has identified that a material change with the printed circuit board assembly (PCBA) has caused performance issues and a second issue with cable connectivity with the speed control dial. In December 20, 2024 Permobil issued a recall the speed control dial potentiometer had a material changes in the potentiometer that resulted in voluntary recall due to allegations of continued drive and involuntary movement. (Health Canada Recall RA-76773). Recall start date: August 12, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.