Merit basixCOMPAK™, PhD™ Hemostasis Valve, and basixSKY™ Inflation Syringe

Health Canada — Medical Devices · 2025-08-14 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-08-14
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Merit Medical became aware of a manufacturing defect after an increased number of complaints. Investigation into the issue concluded that the lack of fluorinated silicone in the bleed back seal (the secondary valve) caused by the mix of incorrect silicone during production line influenced the performance of the PhD™ Hemostasis Valve. The lack of silicone was found to prevent the secondary valve from releasing and closing properly. The valve cap potentially would remain depressed causing leakage. Recall start date: July 30, 2025