Official recall record
Merit basixCOMPAK™, PhD™ Hemostasis Valve, and basixSKY™ Inflation Syringe
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-08-14
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Merit Medical became aware of a manufacturing defect after an increased number of complaints. Investigation into the issue concluded that the lack of fluorinated silicone in the bleed back seal (the secondary valve) caused by the mix of incorrect silicone during production line influenced the performance of the PhD™ Hemostasis Valve. The lack of silicone was found to prevent the secondary valve from releasing and closing properly. The valve cap potentially would remain depressed causing leakage. Recall start date: July 30, 2025
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.