Allura and Azurion Systems
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-08-15
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
The 1 Phase UPS is an optional component that can be installed in Allura and Azurion systems to support a controlled shutdown of the PCs (the computers that control the system) in the event of a mains power failure. The 1 Phase UPS provides temporary power to the PCs, allowing the system to save the data of the last acquired run before the PCs shut down. Philips has become aware of a potential safety issue with Philips Azurion and Allura systems configured with a 1 Phase uninterruptible power supply (UPS). Philips has identified that some internal components of the 1 Phase UPS may fail. When this occurs, there will be a complete loss of power to the system, causing the system to shut down and/or not start up. The 1 Phase UPS system may fail without warning and prior to any mains power failure. If the system does not start up or the system shuts down, there is a potential risk of delay or termination of the procedure. The potential delay in treatment and/or termination of the procedure may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with critical patients, particularly those undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, st-segment elevation myocardial ischemia). Recall start date: July 31, 2025
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.