HistoCore PELORIS 3 (ref 77993)

Health Canada — Medical Devices · 2025-08-27 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-08-27
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Leica Biosystems has become aware of a reagent leakage issue associated with a tubing connection to a manifold in a small number of the HistoCore PELORIS 3 instruments. Our investigation has identified that the root cause of the leakage is due to isolated instances of incorrectly assembled fittings on a batch of manifolds. Leakage may expose instrument operators and other laboratory personnel to toxic reagents and pose a slip hazard. In case of significant leakage, tissue processing may be impacted. Recall start date: August 15, 2025