Official recall record
Getinge Washer/Disinfector
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-08-28
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
The issue found can cause the device to overheat during operation if the circulation pressure calibration was not performed or been performed incorrect and the low-pressure alarm is not triggered. The scope of this action is the electrical heated 88-5 washer disinfectors. Justification: the construction on steam heated devices is different and the heating elements cannot be overheated. This field action is initiated to verify that the circulation pressure in standby mode reads 0.0 kPa. If the value deviates from 0.0 kPa, the device must be recalibrated by a qualified technician to ensure safe continued clinical use. Recall start date: August 21, 2025
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.