Official recall record
Flex Monoclonal Mouse Anti-Human CD20CY, Clone L26, Ready-To-Use (Dako Omnis)
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-09-03
- Last updated (per notice)
- 2025-09-05
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Weak staining has been observed in a few cases where GA604 was used to stain CLL/SLL tissue samples for CD20. This weak staining has not been observed in the majority of cases where GA604 was used to test CLL/SLL tissue. As a reminder, the diagnosis of CLL/SLL is made in the context of the histologic appearance by hematoxylin and eosin (H&E) staining, with the aid of a panel of immunohistochemical markers (routinely including other B cell markers such as PAX 5 or CD79), ancillary studies (e.g. flow cytometry on peripheral blood or bone marrow), and clinical information. Recall start date: July 28, 2025
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.