CellMek SPS

Health Canada — Medical Devices · 2025-09-04 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-09-04
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Issue #1- The instructions for use states that the dead volume for tubes in cassette B is 750ul and cassette E is 1300ul. However, testing shows that the dead volumes are actually 1582ul and 2093ul. Issue # 2- Cassette type E is not rocked prior to aspirating the specimen. Recall start date: August 27, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.