ICEfx™ Cryoablation Console
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-09-11
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Boston Scientific has identified that certain desiccant tube subassemblies, a component of the preventative maintenance (PM) kits used to service ICEfx™ Cryoablation System, were built with end caps that were not correctly tightened. Loose end caps could allow the desiccant to absorb excess atmospheric moisture during storage, potentially impairing argon gas drying and restricting flow due to needle clogging, which could alter therapeutic response. To date Boston Scientific has not received any complaints or reports of patient harm associated with this issue. Recall start date: September 18, 2025