ProductGuru Recalls

Official recall record

HeartSine® Pad-Pak™

Health Canada — Medical Devices · 2025-09-11 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2025-09-11
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

Post-market surveillance has revealed that the Pad-Paks are not always inserted properly into the HeartSine Samaritan® PAD devices as outlined in the instructions for use (IFU), which can create failure during device use. Upon investigation, the potential causes of the improper insertion of Pad-Paks include use error and bent locator pins, which may occur during the manufacturing process. In the event the device is unable to complete connection, the device will repeatedly prompt "apply pads to patient's bare chest". In some cases, the AED device may fail to power on entirely. Recall start date: August 22, 2025

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.