ProductGuru Recalls

Official recall record

Hollow Fiber Hemodialyzers

Health Canada — Medical Devices · 2025-09-16 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-09-16
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

The dialyzer caps had been changed from a threaded version to a push on version whereby the user must push the caps firmly onto the dialyzer to securely affix the cap. As users were accustomed to twisting the caps on, there were complaints of caps coming off resulting in dialysate leaks. two (2) of the complaints received in the United States reported a serious injury to the user related to a slip/fall resulting from the fluid leak. As part of CAPA actions by the manufacturer, IFU (F50010927 rev. B) was updated to IFU (F50010927 rev. C) to specify the method for securing the blood port cap to the dialysate port. As part of this recall, the updated IFU instructions will be communicated.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.