TactiFlex™ Ablation Catheter, Sensor Enabled™

Health Canada — Medical Devices · 2025-09-17 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-09-17
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Abbott is issuing this recall letter related to the TactiFlex™ Ablation Catheter, Sensor Enabled™. As part of their postmarket surveillance and monitoring processes, Abbott became aware of a limited number of tip detachment events during catheter removal from the packaging. Although the packaging has not changed since the introduction of the device (2022), improper removal from the packaging can result in tip damage. While most damaged tips (26 out of 29 instances) were detected prior to catheter insertion into patients, Abbott has experienced 3 recent instances where tips have detached and remained in the patient (incidence rate of <1:100,000). In all three (3) cases there were no further adverse outcomes reported to the patients. Abbott has made changes to the packaging tray design to help mitigate tip damage during catheter removal and has recently started to deploy the new packaging.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.