ProductGuru Recalls

Official recall record

Power X2 Powered Ambulance Cot

Health Canada — Medical Devices · 2025-09-18 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2025-09-18
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

Ferno has identified an issue wherein the fastening post hardware could loosen allowing the fastening post to detach from the cot. The fastening post is intended to secure the ambulance cot in the ambulance fastener during routine patient transport. An unsecured ambulance cot could cause injury to patient and or medic.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.