Official recall record
Ziptight, Acute Ac Joint Implant, Single Ziploop
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-10-07
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Zimmer Biomet is conducting a medical device recall for one lot of the Ziptight, Acute AC Joint Implant, Single Ziploop. Seven complaints have been received with no patient impact, reporting that the slotted button assembly was missing. The defectiveness was discovered on July 15, 2025. Issue was identified through the complaints handling process.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.