Zenition 50

Health Canada — Medical Devices · 2025-10-15 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-10-15
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Philips has been notified by a supplier about an issue affecting integrated chips used in the image intensifier television (IITV) of the Zenition 50 systems. Due to contamination in the bonding layer of the chips on the PCBA (printed circuit board assembly) there could be corrosion of the bonding layer, potentially resulting in intermittent or permanent connection loss of the integrated chips. The issue may result in loss of imaging functionality, or clinically unusable images due to reduced image quality (blurred, reduced image size).

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.