Official recall record
MAS™ Omni-CARDIO™
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-10-16
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Microgenics initiated recall of MAS Omni•CARDIO liquid assayed integrated cardiac control due to reported complaints of vial-to-vial variability when using the control with high sensitivity troponin I assays only (but not with standard/regular troponin I assays). No other analytes are reported to be showing variability between vials. This issue could result in a delay in reporting patient results. If the variability of the material exceeds your lab's specification, dispose of the material.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.