ProductGuru Recalls

Official recall record

HeartMate II® LVAS - System Controller

Health Canada — Medical Devices · 2025-10-17 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2025-10-17
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

Abbott is initiating a voluntary medical device recall for the HeartMate II® and HeartMate 3™ System Controllers. This recall does not involve device removal from where they are used or sold. Abbott has observed an increase in customer complaints about the controller Backup Battery Fault Alarm. This alarm is indicated by a flashing yellow wrench on the user interface of the System Controllers. When the controller is connected to an external power source (e.g. 14V batteries or AC power via the Mobile Power Unit), this is an advisory alarm and does not indicate a critical system failure or impact pump functionality. To reduce future instances of the Backup Battery Fault Alarm, Abbott is advising that when performing backup battery installation or replacement, clinicians should minimize excessive handling, movement, stress and pulling at the interface between the ribbon cable and controller. An example of excessive handling would be using the ribbon cable to hold the weight of the controller. Please remind patients and caregivers to follow the patient handbook: if System Controller Backup Battery Fault Alarm is active, patients should first ensure that no other alarms are active, and then they should call their hospital contact as soon as possible for diagnosis and instructions. Patients and caregivers should not immediately attempt to exchange the system controller without clinical guidance.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.