Official recall record
Catalyst Hd and Catalyst+ Hd Systems
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-10-21
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
C-RAD has noted a potential issue affecting the CSRs mode of the Catalyst+ HD and Catalyst HD systems. The issue is that the stated precision of the catalyst system for stereotactic radiosurgery (SRS) treatments cannot be guaranteed for all couch angles and system setups. in certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance. This could in the worstcase result in erroneous dose delivery.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.