Catalyst Hd and Catalyst+ Hd Systems

Health Canada — Medical Devices · 2025-10-21 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2025-10-21
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

C-RAD has noted a potential issue affecting the CSRs mode of the Catalyst+ HD and Catalyst HD systems. The issue is that the stated precision of the catalyst system for stereotactic radiosurgery (SRS) treatments cannot be guaranteed for all couch angles and system setups. in certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance. This could in the worstcase result in erroneous dose delivery.