What the authority reported

The operator table (part number 14467741) is not intended and not released for bus installations as it was designed for stationary operation only. Due to the impacts and/or vibrations during transportation, there is a risk that parts of the operator table (e.g. screws, monitor, monitor holder) could come loose and potentially fall off resulting in a risk of serious injury to the operating staff from falling parts. In addition, damage to the lead shield may also occur.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →