Official recall record
Xpert® CT/NG
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-10-24
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Cepheid has become aware of a rare genetic mutation event between N eisseria gonorrhoeae and N eisseria meningitidis that caused some N eisseria gonorrhoeae strains to lack target sites for the Cepheid Xpert® CT/NG test, leading to false negative results. Cepheid recommends that laboratories and clinics using the Xpert® CT/NG test as their molecular method for the detection of N . gonorrhoeae monitor for decreases in the positivity rate and ensure that discordant laboratory results (e.g. testing negative on the Xpert® CT/NG test but positive for NG by culture/microscopy or testing negative on Xpert® CT/NG test but positive for NG using another NAAT) be investigated further and reported to cepheid.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.