Digital Diagnostic Mobile X-Ray System GM85

Health Canada — Medical Devices · 2025-10-28 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-10-28
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

We have received a report regarding the loosening or breakage of bolts that connect the tube head unit (THU) to the arm. An investigation determined that this may be caused by metal fatigue resulting from accumulated stress over a long period of use. If all connecting bolts were to fail, it could potentially cause the THU to detach from the arm, posing a risk of serious injury to a patient or user.