ProductGuru Recalls

Official recall record

BD Alaris™ Pump Module

Health Canada — Medical Devices · 2025-10-29 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-10-29
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

BD was made aware via customer complaints that dropping the BD Alaris™ Pump Module may cause damage to internal components that may not be apparent or readily visible. More specifically, drops or severe jarring may damage the BD Alaris™ Pump Module bezel assembly, which provides the mechanical foundation for critical pumping components. This damage can cause under-infusion, over-infusion, unregulated flow or failure of the BD Alaris™ Pump Module to calibrate.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.