Vantage Galan 3T

Health Canada — Medical Devices · 2025-11-06 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-11-06
Last updated (per notice)
2025-11-12
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

It has been found that if there is slow leakage of refrigerant gas from the superconducting magnet over a long period of time, air ingress into the helium vessel via the leak point may result in ice forming from moisture in the air. In the worst case, if the ice is not eliminated, the helium tank could rupture, resulting in a helium leak.