Official recall record
RADspeed Pro X-Ray and CH-200, CH-200M Ceiling Tube Supports
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-11-13
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
The CH-200 and CH-200M ceiling tube supports include a suspension mechanism and rotating shaft that allow x-ray tube positioning. It was found that the mounting shaft may crack and eventually break due to material aging. in rare cases, the protective components may not function properly if the shaft fails while the unit is lifted forward.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.