ProductGuru Recalls

Official recall record

RADspeed Pro X-Ray and CH-200, CH-200M Ceiling Tube Supports

Health Canada — Medical Devices · 2025-11-13 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-11-13
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

The CH-200 and CH-200M ceiling tube supports include a suspension mechanism and rotating shaft that allow x-ray tube positioning. It was found that the mounting shaft may crack and eventually break due to material aging. in rare cases, the protective components may not function properly if the shaft fails while the unit is lifted forward.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.