Tempus Pro
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-12-09
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
The user/operator manual references the term 'invasive pressure', which includes intracranial pressure (ICP) and bladder pressure (BDR) as selectable channel labels. However, these measurements on the Tempus Pro monitor have not been tested or validated for accuracy. The device is only cleared for invasive blood pressure (IBP) measurements.