Beckman Coulter AU Immunochemistry System- HDL/LDL Cholesterol Assay

Health Canada — Medical Devices · 2025-12-10 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-12-10
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

LDL cholesterol OSR6x96 does not meet the triglyceride interference claim per the BAOSR6x96 v10 IFU which states "no significant interference up to 1,500 mg/dl triglyceride" where the criteria for no significant interference is recovery within 10% of the initial value as outlined in the IFU.