ProductGuru Recalls

Official recall record

Edan Patient Monitors

Health Canada — Medical Devices · 2025-12-23 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-12-23
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Labeling issue: an inappropriate publicly routable IP address was used as an example in IFU for demonstrating user-configurable IP address settings. Default setting issue: an inappropriate publicly routable IP address was used as the factory default configuration. However, this IP address is not hard-coded and can be modified by users.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters