Official recall record
Asserachrom® HPIA
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2025-12-23
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
After receiving three customer complaints (from outside Canada and the US), Stago investigated and confirmed that approximately 5% of reagent 1 strips had a defect for Asserachrom® HPIA lot 271288. When the defect is present, every well in the impacted strip produces negative results. Kit reference reagent 6 and control 2 (reagent 7b) produce low optical density (OD) readings, similar to those of control 1 (reagent 7a). As a result, these values will not meet the requirements specified in the assigned value sheet for the lot and must not be used. consignee actions: Stop using lot 271288 and dispose of all remaining product according to local regulations. Complete, sign, and return acknowledgement form.