Asserachrom® HPIA

Health Canada — Medical Devices · 2025-12-23 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2025-12-23
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

After receiving three customer complaints (from outside Canada and the US), Stago investigated and confirmed that approximately 5% of reagent 1 strips had a defect for Asserachrom® HPIA lot 271288. When the defect is present, every well in the impacted strip produces negative results. Kit reference reagent 6 and control 2 (reagent 7b) produce low optical density (OD) readings, similar to those of control 1 (reagent 7a). As a result, these values will not meet the requirements specified in the assigned value sheet for the lot and must not be used. consignee actions: Stop using lot 271288 and dispose of all remaining product according to local regulations. Complete, sign, and return acknowledgement form.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.