Hickman™ and Broviac™ Central Venous Catheter Kits

Health Canada — Medical Devices · 2026-01-20 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-01-20
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

BD has identified via internal inspections that the packaging of some Hickman catheter and Broviac catheter kits may have damage present on the outer tray, potentially compromising the sterile barrier. This product features a double-tray packaging structure (outer tray and inner tray). The inner tray, which contains the catheter and accessories, is sealed inside the outer tray along with the instructions for use. The sterility of the outer surface of the inner tray and the instructions for use may be compromised.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →