What the authority reported

A rare software malfunction has been identified on the cobas® pro (c503) and cobas® pure (c303) platforms affecting the calibration library. This defect only occurs with cobas® pro software version <03-01 or cobas® pure software version <01-04. It can impact assays using spline-type calibration.