Official recall record
Alphenix Systems
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2026-02-10
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
An internal error would occur if F-STORE or image acquisition is performed while CAAS clinical analysis function is used. The manufacturer confirmed that this is caused by a software problem and will not occur if F-STORE or image acquisition are performed after terminating CAAS clinical analysis function.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.