ProductGuru Recalls

Official recall record

Trilogy EVO and EV300

Health Canada — Medical Devices · 2026-02-13 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2026-02-13
Last updated (per notice)
2026-02-18
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

Philips Respironics prohibits the use of non-pneumatic (e.g., vibrating mesh) nebulizers (FSN 2025-CC-SRC-020) with Trilogy EVO platform ventilators (including Trilogy EVO, Trilogy EVO O2, and Trilogy EV300 ventilators). The firm notifies clients about revisions to a prior field safety notice concerning nebulized aerosol deposition on the internal flow sensor (FSN 2024-CC-SRC-013 - flow sensor nebulized aerosol deposition) Philips Respironics is no longer permitting the use of non-pneumatic nebulizers with Trilogy EVO platform ventilators. These updates are necessary for ongoing safe and compliant functioning. Philips Respironics has issued a follow-up letter to Field Safety Notice (FSN) 2026‑CC‑SRC‑002, to provide updated guidance regarding nebulizer use with the Trilogy Evo Platform ventilators (Trilogy Evo, Trilogy Evo O₂, Trilogy EV300). *Updated Guidance: Non‑pneumatic nebulizers may continue to be used, provided that: they are positioned according to the updated nebulizer placement instructions, and the warnings in the revised FSN are fully understood and followed. This update clarifies that non‑pneumatic nebulizers (e.g., vibrating mesh) may continue to be used with Trilogy Evo ventilators, if they are positioned correctly according to the updated placement instructions. This revised guidance supports continuity of care in situations where other nebulization options are limited. Philips Respironics will update the User Manual to include the new guidance on using non‑pneumatic nebulizers. A software update will also be released so the device interface clearly reflects that the guidance applies to both pneumatic and non‑pneumatic nebulizers. These updates will be included in a future release to limit repeated changes for users. NOTE: This update does not change the previously communicated information regarding: • Nebulized aerosol deposition on the internal flow sensor (FSN 2024-CC-SRC-013 – Flow Sensor Nebulized Aerosol Deposition) For further support, contact your local Philips representative.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.