Monarch Platform

Health Canada — Medical Devices · 2026-02-19 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-02-19
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Auris Health, Inc. has identified through internal testing that, in rare cases, a restart of a Monarch software application may occur during procedure setup. If the application restarts after patient-side selection and alignment, but prior to bronchoscope loading, the application will re-initialize the patient-side state to the patient-left position. If the user had previously selected the patient-right position, this re-initialization could result in a mismatch between the system state and the user's intended selection. If unrecognized, this condition could lead to unintended robotic arm movement toward the patient-left position instead of the intended patient-right position during subsequent steps.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.