ProductGuru Recalls

Official recall record

BD Alaris™ Syringe Pumps

Health Canada — Medical Devices · 2026-02-20 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-02-20
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

BD has identified through an internal investigation that the BD Alaris™ syringe pumps may develop a mechanical anomaly on the linear potentiometer if the device is not stored in the correct position. The syringe plunger moves along the linear potentiometer component during infusion. This is an internal component and not readily visible to the clinical user. If this anomaly occurs, it can lead to incorrect sensor readings, which may trigger error codes PL3, PL8, or PL9. These PL error codes are an intentional design feature that notify the clinical user that the syringe pump has detected an internal malfunction and can occur for a number of reasons. The PL3/PL8/PL9 error codes result in the syringe pump issuing an error message along with a red illuminated beacon, and a high priority alarm tone. If an infusion is in progress, it will stop. The error code will be displayed on the syringe pump as <error code and message>, see below image example of a PL9 error code.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.