Automated Impella Controller
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2026-02-25
- Country of alert
- Canada
- Alert level
- Type I
- Alert type
- recall
What the authority reported
Abiomed has identified that Automated Impella Controller (AICS) with software V8.2.3, when used in conjunction with an Impella 5.5 with smart assist set (product code 0550-0004), has a delayed "purge system blocked" alarm. Software versions post-v10.1 have this issue addressed. Abiomed recommends that you do not use the AIC with an Impella 5.5 with smart assist set (product code 0550-0004) unless AIC software is updated to v10.1 or above.
What to do