Automated Impella Controller

Health Canada — Medical Devices · 2026-02-25 · CanadaType I
Issuing authority
Health Canada — Medical Devices
Published
2026-02-25
Country of alert
Canada
Alert level
Type I
Alert type
recall

What the authority reported

Abiomed has identified that Automated Impella Controller (AICS) with software V8.2.3, when used in conjunction with an Impella 5.5 with smart assist set (product code 0550-0004), has a delayed "purge system blocked" alarm. Software versions post-v10.1 have this issue addressed. Abiomed recommends that you do not use the AIC with an Impella 5.5 with smart assist set (product code 0550-0004) unless AIC software is updated to v10.1 or above.