Official recall record
Arjo Tenor Mobile Lift
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2026-03-13
- Last updated (per notice)
- 2026-03-19
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
A supplier‑related brake component defect in the Tenor lift actuator may cause loss of self‑locking and sudden uncontrolled descent under load. All affected units must be removed from use immediately and remain out of service until actuator replacement is completed by authorized Arjo service personnel.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.