ProductGuru Recalls

Official recall record

Arjo Tenor Mobile Lift

Health Canada — Medical Devices · 2026-03-13 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-03-13
Last updated (per notice)
2026-03-19
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

A supplier‑related brake component defect in the Tenor lift actuator may cause loss of self‑locking and sudden uncontrolled descent under load. All affected units must be removed from use immediately and remain out of service until actuator replacement is completed by authorized Arjo service personnel.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters