ProductGuru Recalls

Official recall record

Instinct® Plus Endoscopic Clipping Device

Health Canada — Medical Devices · 2026-03-13 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-03-13
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Cook Medical has received 486 complaints of a malfunction that occurs when the user attempts to open the clip jaws by actuating the handle. The malfunction that may occur is the clip housing detaching from the catheter attachment and the clip remaining attached to the drive wire. The clip will extend from the catheter, but it will remain attached to the internal drive wire instead of opening the clip jaws. The clip cannot be opened if this occurs. Corrective actions have been implemented to mitigate the risk of this failure mode. This action is being taken to withdraw products manufactured prior to the implementation of the corrective actions.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.