Official recall record
Allura Xper and Azurion Systems
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2026-03-17
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch due to: cable or connector damage (internal and/or external) internal microswitch failure (each foot switch pedal activates two independent microswitches that - when both activated - release X-ray) foot switch connector issues resulting from missing or incorrectly applied strain relief No or intermittent X-ray could potentially result in a delay or abort of procedure.
What to do