Official recall record
Soft-Vu and Accu-Vu 4 FR Angiographic Catheters
- Issuing authority
- Health Canada — Medical Devices
- Published
- 2026-03-18
- Country of alert
- Canada
- Alert level
- Type II
- Alert type
- recall
What the authority reported
AngioDynamics, Inc., is conducting a voluntary medical device field safety corrective action or recall to the end-user level based on the non-conformance of a molded hub in specific models and lots of Soft-Vu and Accu-Vu 4 fr angiographic catheters. The affected Soft-Vu and Accu-Vu 4 fr angiographic catheters being recalled (removed) contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub. AngioDynamics has become aware of the non-conformance by way of multiple complaints associated with the affected angiographic catheter devices; however, there have been no complaints or events that occurred in Canada. There are no reports of patient injury as a result of this issue.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.