UniCel DxI 600 Access Immunoassay System-Total T4 Assay

Health Canada — Medical Devices · 2026-03-19 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-03-19
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Access Total T4 calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. In the worst-case scenario, there is a medium safety risk associated with this issue due to the potential to cause false low patient results.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →