AneurysmFlow

Health Canada — Medical Devices · 2026-04-02 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-04-02
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Philips has determined that the mafa ratio does not provide reliable prognostic information regarding aneurysm occlusion following flow diverter stent treatment. the instructions for use (IFU) of AneurysmFlow state that the mafa ratio should not be used for making clinical decisions (section 9.2 measuring flow). In addition, when hovering over the mafa ratio header within the software interface, an on-screen message states that the mafa ratio is not to be used for clinical decision making. despite the existing cautions, the displayed mafa ratio may still be considered during intraprocedural decision-making and could influence clinical judgment. This may potentially result in an incorrect clinical decision.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.