Olympus Connecting Tubes

Health Canada — Medical Devices · 2026-04-02 · CanadaType II
Issuing authority
Health Canada — Medical Devices
Published
2026-04-02
Country of alert
Canada
Alert level
Type II
Alert type
recall

What the authority reported

Olympus identified through complaint review that version 2 connecting tubes may fail prematurely, with lock levers detaching or breaking earlier than expected. The current design resolves issues seen in versions 1 and 2. Olympus will remove prior versions from the market and replace them with the current version.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: Health Canada — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: Health Canada — Medical Devices · Canada · all filters